Cameras for Surface Defect Inspection of Precision Optics

November 23, 2026

Every lens, window, mirror and filter that leaves an optics shop has a cosmetic surface specification: how many scratches and digs it may carry, and how large. For decades that check has been done by trained inspectors holding parts under a lamp against a dark background. It works, but it is slow, it varies from one inspector to the next, and it leaves no image behind.

More optics makers, defense primes and inspection system builders now want a camera to do that job, or at least to record it. This post is for engineers building automated surface imperfection inspection, whether it's a benchtop station in an optics shop or a module inside a larger inspection system. It covers what scratch-dig work asks of a cooled scientific camera, using Finger Lakes Instrumentation (FLI) cameras as examples. For the broader inspection picture, see our post on cooled cameras for display, solar and optics inspection.

Why automate scratch-dig inspection

Three things push shops toward cameras. Repeatability comes first: two inspectors can disagree on whether a faint scratch passes, and an image settles the argument. Second, traceability: a saved image of each surface, tied to a serial number, is far easier to defend to a customer than a checked box. Third, throughput: large optics and high-volume windows take a long time to scan by eye.

What the standards ask for

The two specifications most buyers meet are MIL-PRF-13830B and ISO 10110-7.

MIL-PRF-13830B uses the familiar scratch-dig notation, such as 60-40. The dig number is a size: the diameter of the dig in hundredths of a millimeter, so a dig 40 is 0.40 mm across. The scratch number is different. It is graded by visual comparison with reference standards, not by measuring width. That makes scratch grading the hardest part to automate, because a camera system has to be calibrated against how those standards look under defined lighting.

ISO 10110-7 takes a dimensional approach. A callout such as 5/3 × 0.16 allows three imperfections of grade 0.16, where the grade is the side, in millimeters, of a square with the same area as the defect. Because the grade is an area, a calibrated camera can measure it more directly. ISO 14997 then describes test methods for surface imperfections, both visual and instrument-based.

Before you choose a camera, decide which standard your parts are inspected to. It sets what you must measure, and how small.

Dark-field imaging and weak scatter

Most automated systems copy what the human inspector does: they light the surface at an angle and image it against a dark field, so a clean surface looks black and every scratch or dig shows up as scattered light.

That scattered light is faint. Bennett and Porteus showed that total integrated scatter from a smooth surface scales roughly as (4πσ/λ)², where σ is the rms roughness. On a well-polished optic, the background scatter is tiny, and a small dig only adds a little more. To see it, you need a camera with a low noise floor and low dark signal, so faint defects aren't buried in the camera's own noise during longer exposures. Cooling matters here. Our post on long-exposure dark current in cooled CMOS cameras covers that arithmetic.

Dynamic range: glints next to faint digs

Dark-field images are rarely all dim. The edge of a lens, a chamfer, a mount or a large scratch can throw a bright glint into the same frame as a faint dig. If the camera saturates on the glint, you lose the pixels around it. If you shorten the exposure to stop saturation, the faint defects vanish.

This is where dynamic range matters more than resolution. Kepler cameras build a 16-bit image from two 12-bit readouts at different gains, with less than 1% non-linearity, so bright and faint features can be measured in one exposure. Linearity also matters for grading: if defect brightness is part of your classification, the camera's response has to be predictable across the range. Our acceptance testing guide shows how to check gain and linearity on arrival.

Field of view and pixels per defect

Detection and measurement need different sampling. To detect a dig, a few pixels across it are enough. To measure its size reliably against a grade limit, you want more.

Here is an illustrative example. At 1x magnification, the 3.76 µm pixels of the AR455 cover 3.76 µm on the part, so a dig 10 (0.10 mm) spans about 27 pixels, but the field is only 36 × 24 mm. At 0.25x, each pixel covers about 15 µm and the same sensor sees roughly 144 × 96 mm in one frame. A dig 10 then spans about 6.6 pixels and a dig 5 about 3.3. Scratches narrower than a pixel can still be detected in dark field, because they scatter light, but their width can't be measured at that scale.

Large single frames reduce stage moves and stitching across a big aperture. Larger pixels, such as the 9 µm pixels of the 4096 × 4096 KL4040BI, gather more light per pixel for the faintest scatter. For a compact station, the RV571 puts a 6244 × 4168 sensor in the thin Revelation body. See our post on pixel size and sampling for the general math.

Wavelength, windows and background

Because scatter rises steeply at shorter wavelengths, shifting the illumination from 800 nm to 400 nm makes the same roughness scatter about four times more light. Many systems use blue or near-UV light for that reason. If you do, check that the camera window passes it well. FLI offers window options that reach farther into the near-UV or farther into the near-IR, and the window is chosen before the camera is built in our clean room.

Background control matters as much as the camera. Stray light from the fixture, dust and reflections from the far surface of a transparent part all look like defects. Baffles, clean handling and background frames taken with the part removed help software separate real imperfections from the rest.

Timing, darks and strobed light

Many stations strobe their illumination or step through several lighting angles. Aurora and Revelation cameras have external trigger in and out as standard, so the camera and lights can share one timing source. They support global reset, with rolling readout because that's what the sensor supports. Time the strobe to fall while every row is exposing, and trigger the camera before the light event. Our post on synchronizing cameras with hardware triggers covers the details.

Aurora and Revelation cameras have no mechanical shutter, so dark frames are taken with the illumination off or the light path blocked. Kepler cameras support internal shutters if your process needs them.

Questions to settle before you choose

  • Which standard are your parts inspected to, and what is the smallest defect you must grade?
  • What is the largest clear aperture you need to image, and can you stitch?
  • What illumination wavelength and angles will you use?
  • How bright are the brightest features in the frame compared with the faintest defects?
  • Will you strobe, and how will the camera and lights be triggered?
  • Does the camera need to fit inside an enclosure, and how will heat leave it?

Talk to our sales team

FLI cameras are Made in USA. We machine our housings in-house, build our cameras in a 1,000+ sq ft ISO Class 7 clean room, and test every camera before it ships, with a production test summary by serial number. If you're building a scratch-dig station or an optics inspection system, our sales team can help you match a camera to your parts, optics and lighting. Request a quote to start the conversation.