When a hydrocarbon flame burns, short-lived excited radicals in the reaction zone give off light at their own wavelengths as they relax. The best known are OH near 309 nm in the ultraviolet, CH near 431 nm in the blue and C2 near 516 nm in the green (the Swan band). Imaging that light through narrow filters is one of the simplest ways to see where a flame is releasing heat, and the ratio between bands tracks the local fuel-air mixture. Work from Imperial College London showed that the OH/CH ratio changes steadily with equivalence ratio and that chemiluminescent intensity follows heat release (Combustion and Flame* 139, 2004).
This post is for combustion researchers at universities and national labs, engine and gas turbine test groups, and builders of flame diagnostics instruments. It covers what chemiluminescence imaging asks of a cooled camera, using Finger Lakes Instrumentation (FLI) cameras as examples. For a broader look at cameras in national lab programs, see our post on large-format cameras for national labs.
The camera records light emitted by the flame itself, so no laser is needed. The setup is simple and works through the windows of an optical combustor or engine. The trade-off is that each image is a line-of-sight integral: every pixel adds up light from the whole depth of the flame along that line of view. Turning the images into local, quantitative values takes care in calibration and reconstruction.
Single-shot imaging of turbulent flames with exposures of microseconds usually calls for intensified or high-speed cameras. Many measurements don't need that. Time-averaged images of a steady or statistically steady flame, phase-averaged images of an oscillating flame, multi-band ratio maps and weak emission from lean or diluted flames all reward a cooled, low-noise camera that can integrate for longer and deliver clean, linear pixels. That is where FLI cameras fit.
OH* is often the strongest chemiluminescence band in hydrocarbon flames, and it is the main band in hydrogen flames, but it sits in the ultraviolet. Three parts of the optical path have to pass it:
CH and C2 sit in the visible, where back-illuminated sensors such as the Sony CMOS sensors in FLI Aurora and Revelation cameras collect light well and add very little read noise. A program that needs all three bands may pair a UV-optimized camera for OH with a visible camera for CH and C2*.
Published setups often image OH and CH through band-pass filters about 10 nm wide, with a wider filter for C2*. Narrow filters keep out neighboring bands, but they also cut the signal, which pushes exposures longer.
Sooty flames add a second problem. Hot soot glows across a broad spectrum, and in the visible bands that glow can outshine CH and C2. A common fix is to take a second image through an off-band filter near each emission band and subtract the scaled background. Plan the filter set with those background bands from the start.
Ratio maps such as OH/CH only mean something if both images line up pixel for pixel. With one camera and a filter wheel, geometry stays the same between bands, as long as the flame is steady over the sequence. The FLI High Speed Filter Wheel moves to the adjacent filter in 23 ms on the 6-position model and 30 ms on the 10-position model, which keeps band sequences short. Our post on fast filter switching covers timing, and our post on back focus, flanges and filter wheels covers the mechanical stack. If the bands must be captured at the same instant, two cameras with a dichroic splitter are the usual answer, and then registration becomes a calibration step of its own.
Take dark frames with the flame off or through a blocked filter position; Aurora and Revelation cameras have no mechanical shutter. Flat-field each band, because filter transmission and vignetting vary across the field.
A flame is bright at its core and faint at its edges, and band intensities can differ widely in the same flame. Settle exposure per band so the brightest region stays below saturation, then check that the faint regions still have enough signal. Kepler cameras produce a 16-bit high dynamic range image from two 12-bit readouts at different gains, with less than 1% non-linearity, which helps when one frame holds both. See the Kepler product page.
Long exposures make dark current part of the noise budget, especially through narrow UV filters. Our post on long-exposure dark current shows the arithmetic.
For axisymmetric flames, such as laminar jet and Bunsen flames, Abel inversion turns the line-of-sight projection into a radial distribution of emission. It needs a clean, well-sampled profile, which favors low noise and enough pixels across the flame. Our post on pixel size and sampling covers matching pixels to the optics.
Non-symmetric flames need several views. Chemiluminescence tomography reconstructs the 3D flame from many cameras looking from different angles, and every camera has to be triggered together. External trigger in and out are standard on Aurora and Revelation cameras, and our post on synchronizing multiple cooled cameras covers the timing. The thin, fin-less Revelation body helps when many cameras must fit around a burner.
Flames in gas turbine combustors can couple with acoustics and oscillate. Phase-locked imaging triggers the camera at a fixed point in each pressure cycle and averages many cycles, giving a clean image of the flame at each phase. Aurora and Revelation cameras support global reset with rolling readout, so each row's exposure ends at a slightly different time. Keep the exposure well inside the phase window you care about, and account for readout time in the timing.